Messages and commands displayed in "Non-destructive empty-full" control:
Place empty 1 / 5
P
NM
X
DX
T1
2T1
S
Result
Product
On control completion, the software generates a process summary
(see point 26.3 of this user manual) and the completed control is automatically
saved in the balance's database.
Caution:
A template and an example of a report from product control is provided in point
26.7 of this user manual.

26.5. Destructive empty-full and full-empty control mode

In the "destructive" control mode independently on the size of product batch
over 100 pieces, the sample set by the software for testing is always 20 pieces.
The other control assessing criteria are set in accordance with the regulation.
After using the list to select a product with set parameters for destructive control
with determined measuring "lot" (manufacturing unit) and starting a destructive
control, the software displays commands aiding the control process (as in case of
the controls described above).
Caution:
Editing products in the database in described in point 27.2 of this user manual.
Depending on enabled control mode, the sequence of products to be weighed is
either: "empty-full" or "full-empty".
Caution:
Remember to maintain the weighing sequence of products together with the
packages and empty packages. It is necessary to let the software correctly
calculate mass of a load placed in its corresponding packaging.
- A command on process course and the value
of the measuring lot (manufacturing unit)
- Number of completed measurements of full
packages and quantity of all measurements in
a given batch
- Nominal value of a controlled product
- Average mass of a controlled product
- The value of a disqualifying average
- Number of negative T1 errors in a sample
(actual/disqualifying)
- Number of negative 2T1 errors in a sample
(actual/disqualifying)
- Standard deviation
- Result of product control: positive or negative
- Name of the controlled product