14.2.3. Readout

The function enables the user to adjust balance operation to ambient
conditions (filter settings) or individual user needs. Menu <Readout> contains
the following elements:
FILTER
Each measurement signal before being displayed is electronically
processed for the purpose of obtaining corrected parameters
specific for stable measurement result, i.e. ready to read.
Balance user can influence to some extend the range of signal processing
by selecting an appropriate FILTER value. Available options:
o very fast
o fast
o average / normal
o slow
o very slow
While setting the filtering level the user should consider the actual
operating conditions of a balance. In case of very good operating
conditions set the filter to average/normal or fast, in case of rough
conditions set the filter to slow or very slow.
Caution:
in case of precision balances 3Y series the recommended range of filter
is between very fast
in case of analytical balances and microbalances 3Y series the
recommended range of filter is between average/normal
Value release
This parameter determines displaying the stability marker for a
measurement result.
There are 3 available settings of value release parameter
o fast
o fast + reliable /recommended/
o reliable
Caution:
The speed of stability marker occurrence depends on applied filter value and
applied value release
Autozero function
The function is to automatically monitor and correct zero
indication of a balance.
If the function is enabled, the following measurement results are
compared to each other in constant time intervals. If the results differ
less than declared AUTOZERO range, e.g. 1 division, the balance is
automatically zeroed, and the markers of stable indication
precise zero
÷
average/normal;
are displayed.
- 47 -
÷
very slow.
and