Command Understood But Not Accessible At This Moment; Mass Frame; Frame Format 1 2; Marker Space Character Mass Space Unit Cr Lf - RADWAG UYA 3Y User Manual

Balances 3y series
Hide thumbs Also See for UYA 3Y:
Table of Contents

Advertisement

Immediately send measurement result in current measuring unit
Format: SUI CR LF

Accessible responses:

SUI_I CR LF
MASS FORMAT - mass value in current measuring unit is returned
Frame format
1 2 3
4
S U I
stability
Example:
S U I CR LF – command from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command carried out,
immediate response of mass value in current measuring unit
Where: _ - space
Switch on continuous transmission in basic measuring unit
Format: C1 CR LF
Accessible responses:
C1_I CR LF
C1_A CR LF

MASS FRAME

Frame format
1 2
3
stability
S I space
marker
Switch off continuous transmission in basic measuring unit
Format: C0 CR LF
Accessible responses:
C0_I CR LF
C0_A CR LF
- command understood but not accessible at this
moment
immediately
5
space
sign

- command understood but not accessible at this moment

- command understood and in progress

- response is mass value in basic measuring unit
4
5
space character mass space
- command understood but not accessible at this moment
- command understood and in progress
6
7-15
16
mass
space
6
7-15
- 170 -
17 18 19 20 21
unit
16
17 18 19 20 21
unit
CR LF
CR LF