S_I CR LF
S_A CR LF
MASS FRAME
Frame format:
1
2-3
4
stability
S space
marker
Example:
S CR LF – command sent from a computer
S _ A CR LF - command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out, response
is mass value in basic measuring unit.
Immediately send measurement result in basic measuring unit
Format: SI CR LF

Accessible responses:

SI_I CR LF
MASS FRAME
Frame format:
1 2
3
stability
S I space
marker
Example:
S I CR LF – command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command carried out,
immediate response of mass value in basic weighing unit
Immediately send measurement results from all weighing platforms
basic measuring units
Format: SIA CR LF
Accessible responses:
SIA_I CR LF
MASS FRAME „P1" CR LF
MASS FRAME „P2" CR LF
- command understood but not accessible at this
moment
- command understood and in progress
- response is mass value in basic measuring unit
5
6
space character mass space
- command understood but not accessible at this
moment
- response is immediate with mass value in basic
weighing unit
4
5
space character mass space
- command understood but not accessible at
this moment
- immediate response of mass value in basic
measuring unit from all weighing platforms
7-15
16
6
7-15
16
- 168 -
17 18 19 20 21
unit
CR LF
17 18 19 20 21
unit
CR LF