Accessible Responses; Es Cr Lf; Format: Odh Cr Lf; Accessible Response: Dh_Mass Cr Lf - Command Carried Out - RADWAG UYA 3Y User Manual

Balances 3y series
Hide thumbs Also See for UYA 3Y:
Table of Contents

Advertisement

Mass
- 9 characters with right justification
Unit
- 3 characters with right justification
Give value of high checkweighing threshold HI
OUH CR LF
Format:
Accessible response: UH_MASA CR LF – command carried out

Frame format :

1
2
3
U
H
space
Mass
- 9 characters with right justification
Unit
- 3 characters with right justification
User log in
Format: LOGIN Name, Password CR LF (Name and password have to be
inserted as they are set in the balance – small and capital letters)

Accessible responses:

LOGIN OK CR LF
LOGIN ERRROR CR LF - command understood, an error in name or

ES CR LF

User log out
Format: LOGOUT CR LF
Accessible responses:
LOGOUT OK CR LF
ES CR LF
Send all implemented commands
PC CR LF
Format:
Accessible responses:
PC_A_"Z,T,S,SI,SIA,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,
LOGIN,LOGOUT,PC" – command carried out, terminal has sent all
implemented commands.
4-12
13
mass
space
- command understood, new user is logged in
password occurred, log in failed
- command not recognized (error in format)
- command understood, user is logged out
- command not recognized (error in format)
- 172 -
14
15
16
unit
17
18
19
space
CR
LF

Advertisement