Analog Ranges; Analog Range Key - Ametek 5000 Instrument Manual

Moisture analyzer with multi-point analysis
Hide thumbs Also See for 5000:
Table of Contents

Advertisement

View Alarm Levels
Use up and down arrow keys to highlight setpoint to view; press Enter to display cur-
rent value. Press Cancel to return to Alarm menu
Alter Alarm Levels
If maintenance level security is in effect, the maintenance password will be required
to continue. Use up and down arrow keys to highlight setpoint to alter; press Enter to
display current value. Use the number keys to enter new value; press Enter to save the
change and return to the Alarm menu. Note that the units associated with these values
correspond to the data on the signal; refer to Page 4-2: Alarm Flags for the various flags
and associated unit
Setpoints to be viewed and/or altered are:
High Concentration
alarm will be activated
if system value exceeds
setpoint.
System alarm (CAL
or ZERO ERROR)
will be activated if a
system value exceeds

Analog Ranges

Each analyzer has four fully configurable 4 to 20 mA analog output channels. Channel 2,
3, and 4 outputs have one sensitivity range and Channel 1 output has both high and low
ranges. The upper and lower range limits on all outputs can be adjusted for the desired
resolution. Refer to Page 4-2: Analog Flags for information on configuring the analog
outputs to track system values such as moisture concentration, flow rate, and pressure.
The output range on channel 1 can be toggled between low and high manually or au-
tomatically. The auto-range function switches channel 1 from low to high range when
the signal is close to the upper limit of the low range (~97%). Auto-range also switches
channel 1 from high to low range when the signal reaches ~85% of the upper limit
of the low range. Refer to Page 4-4: Flags1 for information on enabling/disabling the
auto-range function.

Analog Range Key

With normal operation screen displayed, press the Analog Range key to call up the
menu. Use up and down arrow keys to scroll through menu items. When desired item
is highlighted, press Enter and proceed as follows:
3-12 5000 Multi Point Moisture Analyzer
View upper and lower setpoints for both concentration alarms.
Alter upper and lower setpoints for both concentration alarms.
Alarm_1_Hi
The upper limit of concentration alarm 1.
Alarm_2_Hi
The upper limit of concentration alarm 2.
Alarm_3_Hi
The upper limit of concentration alarm 3.
Alarm_4_Hi
The upper limit of concentration alarm 4.
Alarm_1_Low The lower limit of concentration alarm 1.
Alarm_2_Low The lower limit of concentration alarm 2.
Alarm_3_Low The lower limit of concentration alarm 3.
Alarm_4_Low The lower limit of concentration alarm 4.
Max_Cal_Err
Maximum Δ between indicated and stored values.
Max_Span
Maximum allowable cell span.
Max_CD
Maximum allowable coefficient D.