Installing the New Valve

1. Carefully, mount the new proportional solenoid valve on the sample mani-
2. Install the reserved adapter fitting on the inlet and outlet on either side
3. Place the sample manifold plate back in the Oven and secure with the four
4. Reconnect the two (2) 1/16-inch fittings on the sample system connections
5. Replace the Sample Oven cover and secure it with the two (2) nuts (Figure
6. Replace the analyzer cover and secure it with the mounting screws.
PN 305804902, Rev A
fold plate using the two (2) flat-head screws.
of the solenoid and then reconnect the 1/16-inch fitting at the inlet and
outlet of the valve, gently pushing the tubing in and securing with the nut.
(4) mounting screws.
bracket in the center of the Oven (Figure 4-2).
4-1).
Maintenance and Troubleshooting | 4-9