4/40 Flat Head
Screw
Lock
Washer
Flat
Washer
In/Out connections
must be horizontal
Figure 7-6.
Removing the
moisture generator
and pressure
transducer.
Note the position of the moisture generator in the bracket and how it
is oriented to the mounting screws and the gas connections.
NOTE
8.
Remove the old generator from the bracket, and set it aside.
Use caution when disconnecting the moisture generator and pressure
!
transducer from the sample system. The tubes are delicate and care
should be taken not to transfer torque to them.
CAUTION

Installing the new moisture generator

1.
Install the new moisture generator on the mounting bracket in the
correct orientation (Figure 7-6) and remount the bracket on the sample
manifold plate.
2.
Carefully reconnect the 1/16 inch fittings (two on the generator and
one on the pressure transducer).
3.
Place the sample manifold plate back in the oven and secure with the
four mounting nuts.
4.
Reconnect all four 1/16 fittings on the bracket in the center of the oven.
5.
Replace the oven lid, and secure it with the two nuts (Figure 7-1).
6.
Replace the analyzer cover and secure it with the mounting screws.
7.
Power up the analyzer. Wait until the boot-up sequence is complete.
Enter the new moisture generator production code in the software
(refer to Chapter 4 for details).