Download Print this page
Abbott StarClose SE Manual

Abbott StarClose SE Manual

Vascular closure system deployment steps

Advertisement

Quick Links

StarClose SE
Vascular Closure System
Deployment Steps
©2011 Abbott. All rights reserved. ANZ 080091 10/11

Advertisement

loading

Summary of Contents for Abbott StarClose SE

  • Page 1 StarClose SE Vascular Closure System Deployment Steps ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 2 Vascular Closure System Clip Applier Deployment Side-arm Button Slot Finger Post Finger Post Safety Release Clip Delivery Access Port Flex-guide Tube Locator Wings Access Port Plunger Finger Post Thumb Advancer Stabilizer Finger Loop ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 3 StarClose SE Vascular Closure System Clip Delivery Tube A. Clip B. Clip Delivery Tube C. Sheath Splitter ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 4 StarClose SE Vascular Closure System Clip Material: Nitinol (Nickel Titanium Alloy) Actual diameter of the closed Clip: 4 mm, Clip thickness: 0.2 mm 4 mm 0.2 mm ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 5 StarClose SE Vascular Closure System Exchange System J Tip Guide Wire Dilator Exchange Sheath Hub Sheath Tube ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 6 Iliac crest Inferior epigastric Common femoral artery artery Inguinal Profunda ligament Superficial Profunda femoral artery artery • Perform femoral angiogram • Re-Prep groin • Perform nick and spread ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 7 Hold the bottom of the Hub with the left hand. Firmly push the Hub into the Clip Applier. • A “CLICK” should be heard. Gently pull on the Hub to confirm engagement. ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 8 3 centimeters. The initial splitting of the Sheath will be visible above the skin level. • A “CLICK” should be heard to confirm Locator Wing deployment. Check that the number “2” is completely visible in the number window. ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 9 “3” can be seen in the number window. • A “CLICK” should be heard to confirm completion. Check that the number “3” is completely visible in the number window. ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 10 While maintaining downward pressure, depress the Deployment Button with the right thumb to deploy the Clip. • A “CLICK” should be heard to confirm deployment. Maintain the downward pressure for 2 to 3 seconds. ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 11 Use sterile gauze; firmly depress the area around the puncture site to express any blood in the tissue tract. • Test for security of the close by having the patient lift their head off the table and cough, and lift their leg against operator’s hand. ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 12 StarClose SE Vascular Closure System Arterial Perspective “CLICK” 1 “CLICK” 2 “CLICK” 3 “CLICK” 4 ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 13 Check that the number “2” on the Plunger exits the number window completely. • Gently push down on the patient’s body with the left hand. Remove the Clip Applier with the right hand without rocking or twisting. ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 14 Slide the Safety Release to collapse the Locator Wings. Check that the number “2” on the Plunger exits the number window completely. • Gently push down on the patient’s body with the left hand. Remove the Clip Applier with the right hand without rocking or twisting. ©2011 Abbott. All rights reserved. ANZ 080091 10/11...
  • Page 15 Fax: +61 2 8879 2805 Fax: +64 9524 6584 StarClose SE is a registered trademark of the Abbott Group of Companies. All drawings are artist's representations only and should not be considered as an engineering drawing or photograph. Photo(s) on file at Abbott Vascular.