CARDIOMEMS
PATIENT DISCHARGE
PROCEDURE
ASSEMBLING A NEW PATIENT
ELECTRONICS SYSTEM (PES)
¨ Insert the power cable into the power adapter.
¨ Attach the included power cord clip onto the power adapter.
¨ Plug the cable into the power connector plug on the back of the
PES so they are connected all the way. Press the connected plugs
securely into the molded groove.
PAIRING THE NEW PES
¨ Power on the PES by pressing the clear button on the back of the unit. The
button will light up blue.
¨ Remove the handheld unit from the storage area on the right side of the
pillow, and completely unwind the cable from its groove on the PES.
¨ After the system starts, it will prompt for the language and country. To
select the language and country, press the Next button until the desired
language or country appears and then press the Select button.
¨ Next you will be prompted to enter the sensor serial number (SN). Select
the Enter button and then type in the SN. The SN should be written on
the Patient ID card that is included in the plastic bag in the PES box.
Remember to give the patient this ID card for their records once
you have paired the unit.
¨ The PES will then download the necessary information for
the patient from the Merlin.net
(PCN) and, once complete, will prompt you to confirm
the information.
¨ When the system is successfully set up, the screen
will display "Welcome, [patient's name]" above the
Start button.
TAKING THE FIRST PES READING
WITH THE PATIENT
¨ Place the pillow under the patient with the
elevated portion of the pillow at the patient's head.
The location of the sensor implant (routinely left
side) should be in the middle of the pillow, which is
where the center of the antenna is located.
¨ Once the patient is relaxed, have them press the green
Start button located on the handheld unit.
¨ If the patient's position is correct on the pillow, you will hear,
Good position on pillow. Stay still.
HF SYSTEM
Patient Care Network

Advertisement

loading

Summary of Contents for Abbott CARDIOMEMS HF

  • Page 1 CARDIOMEMS HF SYSTEM ™ PATIENT DISCHARGE PROCEDURE ASSEMBLING A NEW PATIENT ELECTRONICS SYSTEM (PES) ¨ Insert the power cable into the power adapter. ¨ Attach the included power cord clip onto the power adapter. ¨ Plug the cable into the power connector plug on the back of the PES so they are connected all the way.