| Handling and Control
Chapter 7
|
7-8
Model 3962

Communications Address - RS485

On units that have the RS485 option, direct communication with the Model 1535 alarm
system can be established. Each piece of equipment connected to the 1535 must have a
unique address. An address of 0-24 can be entered for the environmental chamber. A
setting of 0 is an invalid address that the 1535 will ignore. The factory setting for the
RS485 address is 0.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until 485 ADDR XX is displayed in the message center.
3. Press up/down arrow to change the RS485 address.
4. Press Enter to save the RS485 address.
5. Press the Mode key to return to Run mode or right/left to go to next/previous
parameter.
Figure 7-5
Configure Mode
Thermo Scientific