Water Quality Recommendations:
For trouble-free operation, fill the water reservoir with sterilized distilled water or water treated to an
equivalent quality level. The acceptable conductivity should be within the range of 1 to 20 S/cm
(resistivity should be within the range of 50 kOhmcm to 1 MOhmcm).
CAUTION Termination of warranty!
Using chlorinated tap water or additives that contain chlorine will void the manufacture warranty.
Similarly, the use of ultrapure water whose conductivity is out of the range of 1 to 20 S/cm and
whose resistivity is out of the range of 50 kOhmcm to 1 MOhmcm will void the manufacture
warranty.
If you should have any questions, please contact Technical Support of Thermo Fisher Scientific.
CAUTION No tap or ultrapure water in humidity reservoir
Sterile distilled water or water treated to an equivalent quality level is the recommendation for use
in the integrated humidity reservoir. The acceptable conductivity should be within the range of 1 to
20 μS/cm (resistivity should be within the range of 50 K-Ohm-cm to 1 M-Ohm-cm). The pH
should be in the range of 7-9. Ultrapure Type 1 water or deionized (DI) water with a resistivity
approaching or equal to 18.2 M-Ohm-cm, contains very few ions and will actively pull ions from
interior components, damaging stainless steel, copper, and glass. If there is only access to DI or
Type 1 water, one option is to add a sterile solution of weak sodium bicarbonate to raise the pH and
add ions (recommend 84 mg/l (1 mmol/l).
CAUTION No chloride containing disinfectants
While stainless steel is resistant to corrosion, it is not corrosion proof. Many chemicals have a
negative effect on stainless steel, especially chlorine and derivatives with oxidizing activity.
The addition of chloride-containing disinfectants or copper sulfate into the water as a constant
disinfectant is not recommended because these can damage the connecting stainless steel/copper
joint of the drain.For cleaning the interior, the recommendation is mild soap solution in water then
rinse to remove residues. Wipe the interior surfaces and parts with a diluted quaternary ammonium
disinfectant. Follow this by wiping with 70% alcohol to remove any remaining traces of the
disinfectant.
Under normal operating conditions and at the usual incubation temperature of 37 °C, a constant
relative humidity of approx 93% is achieved in the workspace.
If condensation occurs on the culture containers due to high relative humidity, the humidity in the
workspace can be set to a lower level. The relative humidity in the workspace is lowered from approx.
93% to approx. 90%. The modification requires an extended adaption phase. To ensure that it
effectively prevents dew formation on culture containers, it must be used as a permanent setting.
Instructions for activating the Low Humidity function are given in section
page
6-33.
Prefilter
Placed on the front section of the water reservoir cover is a pre-filter. The pre-filter consists of a
dual-layer wire mesh with a silicone frame and is autoclavable as well as heat-resistant. The pre-filter
must remain installed in the device during the Steri-run decontamination routine, but needs to be
removed for refilling the water reservoir.