12
Data Communication
Error Memory Structure
Overview of the Possible Error Messages in Hex Coding
Hex Code
0x00
0x01
0x02
0x07
0x08
Overview of the Possible Error Messages in Bit Coding
General Device Status, Temperature and CO
Bit
0x0002
0x0004
0x0008
0x0010
0x0020
0x0040
0x0080
0x0100
0x0400
0x1000
0x2000
0x4000
0x8000
Bit
0x0001
0x0002
0x0004
0x0008
0x0010
Bit
0x0001
0x0002
HERA
CELL VIOS
12-10
Description / Type
Temperature control loop
CO
2 control loop
O
2 control loop
Water level
General device status
General device status
Device door open too long
Display not communicating
Implausible mainboard parameter (EEPROM defective)
Data logger defective (unit can continue to function)
Error for disinfection / steri-run
Power-down during steri-run
Error auto-start
ADC-test failed
Fan error
IR sensor system replaced (info)
auto-start active (info)
Disinfection active (info)
Device error has occurred (info)
Error status of temperature control circuit
Sensor breakage
Actual value high
Actual value low
Actual value not plausible
Calibration values too high/low
CO
control loop error status
2
Sensor breakage
Actual value high
160i LK / 250i LK
Control Loop:
2
Thermo Scientific