Step 4. Remove the Front Panel Assembly

Refer to
1. With a 5/16-in wrench, remove all front panel jumpers (item
2. With a T-10 TORX driver, remove the screws (item
3. Slide the front panel over the test port connectors.
4. Disconnect the front panel interface ribbon cable (item
Figure 2
Front panel Assembly Removal
Installation Note N5242-90120
Figure 2
for this step of procedure.
the frame.
Before removing the front panel from the analyzer, lift and support the
front of the analyzer chassis.
panel is now free from the analyzer.
).
) from the sides of
). The front