Aespire 7100
9-68
9.23.1
To remove
1. Disconnect pipeline supplies; close cylinder valves; bleed off pressure.
the Vent Englne
2. Remove the ABS breathing system.
A
3.
Remove the
Ex
halation valve.
4. If present
,
remove the scavenging downtube.
5. Loosen the five captive screws (A) that hold the Vent Engine coverto the housing.
Raise the cover to access the Vent Engine.
E
E
B
6. Disconnect the Vent Engine harness (8).
c
E
D
7. Disconnectthe white
tube-coupler(C) 一 inline
with tube to pressure transducer on
the Vent Engine Board.
8.
If present
,
disconnect the black tube-coupler
,
inline with tube to scavenging.
9. Disconnect the drive gas hose (D).
10. Loosen the three captive screws (E) that hold the engine manifold to the housing.
11.
Li ft
the Vent Engine out of the housing
12. To replace the Vent Engine
,
reassemble in reverse order.
08/07 Ml110140