CARESCAPE ONE User Manual DRAFT 26 October 2017
Temperature
3. Select T1, T2 or T2-T1 alarm field.
If the feature is not active, the alarm limits are greyed out. Activate the alarm by
selecting Alarm On. If informational priority level has been selected for the alarm,
the menu selection is Message On.
4. Set the alarm limits.
If the setting of an alarm limit has been disabled during configuration, the setting is
marked with a lock symbol:
Stopping the temperature measurement
1. Select the temperature parameter window.
2. Deselect the measurement check box (e.g., T1 Measurement).
Temperature practicalities
Each temperature label can be changed to reflect the temperature measurement
●
site.
Individual temperature sites can be turned off.
●
The difference between two temperature sites can be calculated and displayed.
●
The CARESCAPE Temperature allows a two-channel measurement.
●
The signal input is a high-insulation port to ensure patient safety and to protect the
●
device during defibrillation and electrosurgery.
The temperature measurements are automatically calibrated at startup and then
●
at least once a minute.
Temperature troubleshooting
Problem
Temperature measurement fails
182
Solution
Check that the CARESCAPE Temperature is
●
properly connected to the acquisition platform.
Check that there is only one CARESCAPE
●
Temperature connected to the acquisition
platform.
Check that the probe is properly connected to the
●
CARESCAPE Temperature or interconnect cable.
Check that you are using the correct probe for the
●
anatomical location being monitored.
Use a probe that is compatible with your system.
●
Try using a known good probe in case the probe is
●
damaged.
If you are using the CARESCAPE ONE with a
●
Bx50 host, check the connection between the
acquisition platform and the host.
Check the patient connection.
●
CARESCAPE ONE
2094480-001