1.3.3 Door Size Requirements

Minimum door sizes also apply to hallways and elevators. For additional details, refer to
Information on page 5
Door Height
The minimum door height (to accommodate Innova positioner on its dolly) is 2001 mm (78.8 in).
Door Width
The minimum door width needed (to accommodate the Innova Frontal/Lateral shipping dolly) is 1040
mm (41 in) with the cable inlet and the dolly stabilizers removed.
NOTE
Door widths are based on a straight–in approach requiring a 2440 mm (96 in) wide
corridor. Calculations need to be made for accommodation of equipment through
narrower corridors.
Elevator Depth
The minimum elevator depth needed to accommodate the Innova Gantry shipping dolly is 2954 mm
(116.3 in).

1.3.4 Route Survey

Step One — Sketch
Start preparing Route Survey by sketching a floor plan of the hospital or clinic which will receive the
equipment. Include all areas on the delivery route from outside the building to destination. See
1-16 on page
24.
Reference Numbers: Numbers in circles refer to Route Survey data. The Route Survey is a form on which
site data are listed (see
5750182-1EN 3
Step Two — Survey on page
TM
Innova
IGS 6
24).
General Requirements
1.3.1 Shipping
Figure
23