Monitor Settings

Patient-Monitor Type

Monitor Defaults Password

Revision A
Configuration: Monitor Settings
WARNING
Changing the patient-monitor type will default the admit
function to Standard configuration. Different alarms and
parameters are activated for each selection.
The patient-monitor type defines the type of monitor desired (i.e adult,
neonatal or operating room). The patient-monitor type displays on the
top left of the display, immediately after the date and time.
To set the patient-monitor type, follow this procedure.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Use the TRIM KNOB control to enter the service password.
3. Select PATIENT-MONITOR TYPE.
4. Select ADULT-ICU, NEONATAL-ICU or OPERATING ROOM.
5. Press the TRIM KNOB control to exit.
The monitor defaults password defines whether or not a password is
required to change the monitor's defaults. If selected, the password will
be the same as the Service Mode menu password.
To define the monitor defaults password option, follow this procedure.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Use the TRIM KNOB control to enter the service password.
3. Select MENU SETUP > MONITOR DEFAULTS PASSWD.
4. Select either REQUIRED or NOT REQUIRED.
5. Press the TRIM KNOB control to exit.
Transport Pro
2012659-042
616A
4-3