Error Identifier Single Plane Measurement; Error Identifier Caustic - Primes FocusParameterMonitor FPM 60 Original Instructions Manual

Table of Contents

Advertisement

8.3.3

Error identifier single plane measurement

Group message "Error Single-Plane Measurement" is set.
„Measurement Failure" is set.
Error
Trigger-Timeout occurred
Error with upstream measurement
Error with raw data measurement
Error with underground measurement
Timeout measurement
Timeout calculation
No beam found with BeamFind
Volume negative
beamdata.r2E < 0 (2. moments)
beamdata.x2E < 0 (2. moments)
beamdata.y2E < 0 (2. moments)
8.3.4

Error identifier caustic

Group message "Error Caustic" is set.
„Measurement Failure" is set.
Error
No focal length presetting
Plane counter: Input incorrect
Too little planes for caustic evaluation
Evaluation error; not calculable
36
Error condition
No laser pulse within 2 s after start.
Problem of FPGA process.
Measurement not executed within 4 s.
Calculations after measurement not executed within 4 s
Possible cause: FTP transfer enabled, but no FTP server found.
Laser was off during measurement time.
Beam parameters couldn't be calculated.
Sensor signal not plausible.
Probably Laser was off during measurement time,
or inappropriate integration time.
Error condition
No focal length transmitted.
Plane number < 0 or >= 30.
Less than 3 planes taken during cauctic measurement.
Caustic fit not successfull.
FocusParameterMonitor FPM
Identification
0x0001
0x0002
0x0004
0x0008
0x0010
0x0020
0x0040
0x0080
0x0100
0x0200
0x0400
0x0800
0x1000
0x2000
0x4000
0x8000
Identification
0x0001
0x0002
0x0004
0x0008
0x0010
0x0020
0x0040
0x0080
0x0100
Revision 06 EN 02/2023