Compressed Air Connection - Primes FocusParameterMonitor FPM 60 Original Instructions Manual

Table of Contents

Advertisement

7.5

Compressed Air Connection

CAUTION
Damage to hearing
If the compressed air hose is loosened while still under pressure, this generates noise.
This can cause hearing damage.
 Only loosen the compressed air hose when there is no longer any pressure.
NOTICE
Risk of damage to optical components
Contaminated compressed air can cause permanent damage to the optical components of the
measuring device
 The compressed air must comply with purity class ISO 8573-1:2010 [1:4:2].
Connecting the compressed air hose
 Push the hose into the plug-in connection as far as possible.
Disconnecting the compressed air hose
 Press down the blue release ring of the connector with one hand and pull out the hose with the other hand.
Data
Outer diameter of the hose
Flow rate
(typical)
Maximum allowable pressure
Purity classes
To save energy, reduce the flow rate to 1...2 l/min when the shutter is closed.
Fig. 7.8:
Device view, compressed air connection
28
(Polyurethan)
FocusParameterMonitor FPM
Value
4 mm
10 - 15 l/min
2 bar
ISO 8573-1:2010 [1:4:2]
Compressed air
connection ∅ 4 mm
Revision 06 EN 02/2023