To Install Ferrules on a Packed Metal Column

C A U T I O N
Maintaining Your GC
1
Gather the following:
Wrenches
Stainless steel male Swagelok fitting, 1/4- or 1/8- inch
od
Brass Swagelok nut and ferrule set (See
and Parts for the Purged Packed Inlet"
Lint- free gloves
2
Verify that the column end is cut square and is free of
burns and deformation.
Secure the fitting in a bench vise.
3
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Assemble a Swagelok nut and ferrules onto the column.
4
Front ferrule
Back ferrule
Column nut
5
Fully insert the column into the vise- held fitting, then
withdraw 1–2 mm. Finger- tighten the nut.
Tighten the nut an additional 3/4 turn with a wrench (for
6
a 1/8- inch column) or an additional 1- 1/4 turn (for a
1/4- inch column).
Unscrew the column nut from the vise- held fitting and
7
remove the column. Ferrules should now be set in place
on the column with the column end correctly positioned.
Maintaining the Purged Packed Inlet
"Consumables
on page 68.)