Agilent Technologies Agilent 7890 Series Maintenance Manual page 183

Gas chromatograph
Table of Contents

Advertisement

Maintaining Your GC
Position the column so it extends 6 mm above the end of
8
the ferrule. Slide the septum up the column to hold the
column nut at this fixed position.
6 mm
Insert the column into the interface and finger- tighten the
9
column nut.
10
Adjust the column (not the septum) position until the
septum is snug against the bottom of the nut.
Tighten the column nut an additional 1/4 to 1/2 turn with
11
a wrench so that the column cannot be pulled from the
fitting with gentle pressure.
Configure the new column. If using the optional barcode
12
scanner, scan the column directly into your data system
software configuration, then download the changes to the
GC.
With the sample transfer line attached and the column
13
attached to the inlet and detector, establish a flow of
carrier gas through the transfer line. Purge as
recommended by the column manufacturer.
14
Condition the column per the manufacturer's
recommendation. (See
Install the column into the detector. See:
15
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the uECD
To install a Capillary Column in the FPD Plus
To Install a Capillary Column Adapter in the FPD
After the column is installed at both inlet and detector,
16
establish a flow of carrier gas and purge as recommended
by the column manufacturer.
Restore the analytical method.
17
For FPD, immediately turn off the flame.
For NPD, immediately set the bead voltage to 0.0.
18
After the GC becomes ready, wait 10 minutes then ignite
the detector flame or adjust offset on the NPD bead.
Maintaining the VI
To Condition a Capillary
10
Column.)
183

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents