4
Maintaining Capillary Columns
To Attach a Capillary Column Using SilTite Metal Fittings
C A U T I O N
40
This procedure is used to attach a capillary column to a
splitter or switch or an Ultimate Union.
Gather the following:
1
•
SilTite ferrules (see
•
Swaging tool (G2855- 60200) or Swaging nut for SilTite
ferrules (G2855- 20555)
•
Two 1/4- inch open- end wrenches
•
One 7/16- inch open- end wrench
•
Column cutting tool (5181- 8836)
•
Internal nut (G2855- 20530)
•
Lint free gloves
Table 5
Available SilTite metal ferrule packages
Part number
5188-5361
5188-5362
5188-5363
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Pass the tubing end through the internal nut and SilTite
2
ferrule leaving approximately 1 cm of fused silica tubing
protruding beyond the ferrule. Thread the swaging tool
over the column.
Table
5)
SilTite ferrule description
For 0.2- to 0.25-mm columns
For 0.32-mm columns
For 0.53-mm columns
Maintaining Your GC