Agilent Technologies Agilent 7890 Series Maintenance Manual page 73

Gas chromatograph
Table of Contents

Advertisement

Maintaining Your GC
Ferrule
Column nut
Septum
Score the column using a glass scribing tool. The score
6
must be square to ensure a clean break.
7
Break off the column end by supporting it against the
column cutter opposite the scribe. Inspect the end with a
magnifying loupe to make certain there are no burrs or
jagged edges.
8
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
Position the column so it extends 1 to 2 mm above the
9
end of the ferrule. Slide the septum up the column to
hold the column nut at this fixed position.
Maintaining the Purged Packed Inlet
Bad
Good
6
73