To Install Capillary Column Clips - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

To Install Capillary Column Clips

WA R N I N G
WA R N I N G
Maintaining Your GC
Be careful! The oven may be hot enough to cause burns. If the
oven is hot, wear heat-resistant gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Gather the following:
1
Capillary column clip kit (see
for Columns"
on page 30.)
T- 20 Torx screwdriver
Loosen the four corner mounting screws, but do not
2
remove.
Column
clip
Mounting
screw
3
Slip each corner screw through the large hole on the clip.
4
Slide the clip so that the screw is positioned in the slot.
5
Tighten the screws enough to hold the clips in place.
Once the column is installed, fully tighten the four corner
screws to secure the clips and column to the oven wall.
Maintaining Capillary Columns
"Consumables and Parts
4
33