Agilent Technologies Agilent 7890 Series Maintenance Manual page 296

Gas chromatograph
Table of Contents

Advertisement

15
Maintaining the FPD
Score column here
Column measuring tool
Column nut
Ferrule
296
+
Insert the end of the column through the column
4
measuring tool so that the end protrudes beyond the tool.
Tighten the column nut until it grips the column. Tighten
5
the nut an additional 1/8 to 1/4 turn with a pair of
wrenches. Snug the septum against the base of the
column nut.
6
Use a wafer cutter at 45° to score the column.
7
Snap off the column end. The column may protrude about
1 mm beyond the end of the tool. Inspect the end with a
magnifying loupe to make certain that there are no burrs
or jagged edges.
Remove the column, nut, and swaged ferrule from the
8
tool.
9
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
Carefully thread the swaged column up into the detector
10
fitting. Finger- tighten the column nut, then use a wrench
to tighten an additional 1/8 turn.
137.5 mm
Bad
Good