To Replace the FID Collector Assembly

C A U T I O N
Maintaining Your GC
Gather the following:
1
New FID collector assembly. (See
Parts for the FID"
T- 20 Torx screwdriver
1/4- inch nut driver
Tweezers
Lint- free gloves
To avoid contaminating the FID, wear clean, lint-free gloves when
handling the collector assembly.
Load the
GC maintenance method
2
become ready.
Disconnect the ignitor cable assembly.
3
Remove the three screws holding the collector assembly to
4
the mounting pallet.
Maintaining the FID
"Consumables and
on page 196.)
and wait for the GC to
11
209