To Install The Column Nut Adapter - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

8
Maintaining the MMI

To Install the Column Nut Adapter

WA R N I N G
132
1
Gather the following (see
MMI"
on page 128):
Column nut adapter
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
When the GC becomes ready, open the GC oven door.
3
Thread the column adapter nut onto the base of the inlet.
4
Column nut adapter
Orient as shown.
The nut must spin freely before it can be used to install a
column.
"Consumables and Parts for the
and wait for the GC to