To Install a Capillary Column with the MMI

WA R N I N G
WA R N I N G
WA R N I N G
C A U T I O N
Maintaining Your GC
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
Gather the following (see
1
MMI"
on page 128):
Column
Ferrule(s)
Column nut
Septum
Column cutter
Isopropanol
Lab tissue
Metric ruler
1/4- inch open- end wrenches
Lint- free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Verify that the correct glass liner is installed. (See
3
"Consumables and Parts for the MMI"
4
Place the column on the hanger with the ends pointing up
and the label to the front.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the MMI
"Consumables and Parts for the
and wait for the GC to
on page 128.)
8
133