To Change the Septum on the MMI

WA R N I N G
Maintaining Your GC
1
Gather the following:
Replacement septum. (See
the MMI"
on page 128.)
Wrench, hex for changing septum
0- or 00- grade steel wool (optional)
Tweezers
Load the
inlet maintenance method
2
to become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Remove the septum retainer nut or Merlin cap.
3
4
Use tweezers to remove the septum or Merlin Microseal
from the insert assembly. Do not gouge or scratch the
interior of the insert assembly.
Septum retainer nut
Septum
Insert assembly
5
Firmly press the new septum or Merlin Microseal into the
fitting. The metal parts side of the Merlin Microseal
should face down (toward the oven).
6
Install the septum retainer nut or Merlin cap and
finger- tighten. Tighten the septum retainer nut until the
C- ring is about 1 mm above the nut.
Maintaining the MMI
"Consumables and Parts for
and wait for the GC
8
137