Maintaining the MMI
134
Place a septum, capillary column nut, and ferrule on the
5
column.
Ferrule
Column nut
Septum
6
Score the column using a glass scribing tool. The score
must be square to ensure a clean break.
7
Break off the column end by supporting it against the
column cutter opposite the scribe. Inspect the end with a
magnifying loupe to make certain there are no burrs or
jagged edges.
Wipe the column walls with a tissue dampened with
8
isopropanol to remove fingerprints and dust.
9
Position the column so it extends 10 to 12 mm above the
end of the ferrule. Slide the septum up the column to
hold the column nut at this position.
Bad
Good