Maintaining Your GC
Silver seal
Graphpak inlet adapter
Column nut
Ferrule for Graphpak adapter
5
Remove the inlet adapter and discard the old silver seal.
6
Insert a new silver seal into the adapter and
finger- tighten the adapter onto the inlet. Tighten an
additional 1/16 to 1/8 turn with a wrench; overtightening
will damage the inlet.
Install the column. (See
7
with the PTV Inlet"
Check the adapter for leaks.
8
9
Restore the analytical method.
10
Reset the silver seal EMF counter. See
Counter
in the Operation Manual.
Maintaining the PTV Inlet
"To Install a Capillary Column
on page 155.)
To Reset an EMF
9
171