14
Maintaining the NPD
284
Install the bead assembly and restore normal operating
26
conditions. (See
"To Replace the NPD Bead Assembly"
page 272.) (Do not reset the bead counter unless
replaced.)
After installing new collector parts, the NPD leakage
current should be lower. (See
Current"
on page 285.) If the leakage current is abnormal,
check for proper reassembly of the detector (especially
where the electrometer interconnect contacts the collector
assembly) and for leaks.
Reset the EMF counters. See
27
the Operation Manual.
"To Check the NPD Leakage
To Reset an EMF Counter
Maintaining Your GC
on
in