Using A Swagelok Tee - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

Using a Swagelok Tee

To supply gas from a single source to more than one input,
use a Swagelok Tee.
Do not combine valve actuator air with flame ionization air. The valve
N O T E
action will cause major upsets in the detector signal.
Materials needed:
• 1/8- inch preconditioned copper tubing
• Tubing cutter
• 1/8- inch Swagelok nuts and front and back ferrules
• 1/8- inch Swagelok Tee
• Two 7/16- inch wrenches
• 1/8- inch Swagelok cap (optional)
Cut the tubing where you want to install the Tee. Connect
1
the tubing and Tee with a Swagelok fitting. See
Cap
Tee fitting
Nut and ferrules
Figure 15 Swagelok tee
2
Measure the distance from the Tee to the GC inlets.
Attach copper tubing to the open Tee ends with Swagelok
fittings.