Agilent Technologies Agilent 7890 Series Maintenance Manual page 253

Gas chromatograph
Table of Contents

Advertisement

13
Maintaining the uECD
If the id is less the 200 um, mark the column with a
septum 70 ± 1 mm from the end. Insert column and nut
into the adapter with the septum at the rear of the
column nut, and tighten the column nut with one
5/16- inch wrench on the adapter and another 1/4- inch
wrench on the column nut.
70 ± 1 mm
Septum
After heating the detector, retighten the 9/16- inch
9
makeup adapter nut and 1/4- inch column nut.
Maintaining Your GC
253