7
Maintaining the COC Inlet
To Replace the 7693A Injector Needle Support Guide
118
Before using a 7693A ALS to make injections onto the COC
inlet, install the on- column needle support guide.
Refer to the
7693A ALS documentation
details.
1
Open the injector door.
2
Remove the syringe.
3
Slide the syringe carriage up to the top position.
4
Completely remove the T- 10 Torx screw from the support
foot. Be careful to not let the screw fall into the turret
assembly.
5
Slide off the support foot.
6
Slide on the new support foot.
7
Replace the T- 10 Torx screw and tighten.
8
Install the appropriate syringe.
9
Close the injector door.
10
Align the injector.
for complete
Standard needle
support guide
On-column needle
support guide
Maintaining Your GC