Agilent Technologies Agilent 7890 Series Maintenance Manual page 109

Gas chromatograph
Table of Contents

Advertisement

WA R N I N G
WA R N I N G
Maintaining Your GC
To avoid bending the inlet, always use two wrenches. Use a
5/16-inch wrench to support the inlet while tightening the column
nut with a 1/4-inch wrench.
Tighten the column nut an additional 1/4 turn with a
10
wrench or until the column does not move.
11
If using an automatic injection system with a 0.25- mm or
0.32- mm column, verify that the column installation by
manually pushing the syringe into the inlet.
12
Configure the new column. If using the optional barcode
scanner, scan the column directly into your data system
software configuration, then download the changes to the
GC.
Condition the column per the manufacturer's
13
recommendation. (See
Install the column into the detector. See:
14
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the uECD
To install a Capillary Column in the FPD Plus
To Install a Capillary Column Adapter in the FPD
15
After the column is installed at both inlet and detector,
establish a flow of carrier gas and purge as recommended
by the column manufacturer.
16
Restore the analytical method.
For FPD, immediately turn off the flame.
For NPD, immediately set the bead voltage to 0.0.
After the GC becomes ready, wait 10 minutes then ignite
17
the detector flame or adjust offset on the NPD bead.
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If they are hot, wear heat-resistant gloves to protect
your hands.
18
Allow the oven, inlet, and detector to equilibrate at
operating temperature, then retighten the fittings.
Reset the column EMF counters. See
19
Counter
in the Operation Manual.
Maintaining the COC Inlet
To Condition a Capillary
To Reset an EMF
7
Column.)
109

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents