Maintaining the FPD
C A U T I O N
294
+
Install the 1/8- inch packed metal column onto the column
4
adapter.
5
Finger- tighten the column, then tighten an additional
1/8 turn using two wrenches.
Handle the adapter carefully and install the packed column before
installing the adapter into the detector base. The thin-walled
capillary tube that carries the sample into the detector can be
damaged by rough handling.
Place the adapter nut and ferrule onto the adapter.
6
Carefully install the adapter assembly into the detector
7
fitting. Align the adapter so the it enters the detector
fitting as vertically as possible. Avoid stressing the
adapter capillary tube. The adapter nut should turn
freely.