Maintaining Your GC
Remove the O- ring with tweezers.
6
Assemble the parts for the new ignitor.
7
Glow plug
Spacer
O-ring
Install the new ignitor assembly and tighten with a
8
wrench. Do not overtighten.
Replace the ignitor collar and cable assembly and tighten
9
the screw.
10
Replace the cover and the vent tube assembly. (See
Install the FPD Vent Tube and Cover"
Restore the analytical method.
11
Wait 20 minutes for the detector to heat up, then ignite
12
the flame.
13
Reset the EMF counter. See
the Operation Manual.