Agilent Technologies Agilent 7890 Series Maintenance Manual page 217

Gas chromatograph
Table of Contents

Advertisement

Maintaining Your GC
Remove the FID ignitor.
3
Disconnect the ignitor cable assembly.
a
Loosen the ignitor with a wrench.
b
c
Turn the nut counterclockwise by hand. Remove the
ignitor and copper washer.
If replacing only the FID ignitor assembly with copper
4
washer, skip to
step 16
5
Remove the three screws that hold the collector mount to
the FID thermal strap.
Maintaining the FID
for assembly.
11
217