Maintaining the COC Inlet
108
6
Break off the column end by supporting it against the
column cutter opposite the scribe. Inspect the end with a
magnifying loupe to make certain there are no burrs or
jagged edges.
7
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
Gently insert the column into the inlet until it bottoms.
8
You should feel spring tension as you push up on the
column. (Do not withdraw the column.)
Insert the column nut into the inlet fitting and
9
finger- tighten.
Bad
Good
Maintaining Your GC