To Install A Capillary Column In The Uecd - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

To Install a Capillary Column in the uECD

WA R N I N G
WA R N I N G
C A U T I O N
Maintaining Your GC
1
Gather the following:
Ferrule (See
"Consumables and Parts for the uECD"
page 244.)
Column nut
Septum
Column
1/4- inch, 5/16- inch, and 9/16- inch wrenches
Column cutter
Lint- free gloves
Load the
GC maintenance method
2
become ready.
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Load the
inlet maintenance method
3
to become ready.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
4
Place a septum (if the column id is
column nut, and ferrule on the column.
Maintaining the uECD
and wait for the GC to
and wait for the GC
0.2 mm), capillary
13
on
251