To Remove the VI Interface

WA R N I N G
Maintaining Your GC
1
Gather the following:
1/4- inch and 7- mm wrench
T- 20 Torx screwdriver
2
Load the
inlet maintenance method
to become ready.
Be careful! The interface may be hot enough to cause burns. If the
interface is hot, wear gloves to protect your hands.
3
Remove the column.
4
Remove the transfer line by loosening the nut, then lift it
from the interface.
Loosen the five inlet cover screws and remove the inlet
5
cover.
6
Remove the clamping plate from the interface by
loosening the captive screw.
Lift the interface out of the heater block.
7
10