To Clean The Split/Splitless Inlet - Agilent Technologies Agilent 7890 Series Maintenance Manual

Gas chromatograph
Table of Contents

Advertisement

5
Maintaining the Split/Splitless Inlet

To Clean the Split/Splitless Inlet

WA R N I N G
C A U T I O N
64
1
Gather the following:
Replacement septum (See
the Split/Splitless Inlet"
Replacement liner
Replacement O- ring
Replacement gold seal
Replacement washer
Solvent that will clean the type of deposits in your
inlet
Compressed, filtered, dry air or nitrogen
Beaker
Cleaning brushes—The FID cleaning kit (part number
9301- 0985) contains appropriate brushes
Lint- free gloves
Load the
inlet maintenance method
2
to become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Remove the inlet liner. (See
3
O- Ring on the Split/Splitless Inlet"
4
Disconnect the column from the inlet.
5
Remove the reducing nut and gold seal. (See
the Gold Seal on the Split/Splitless Inlet"
Place a beaker in the oven under the inlet to catch the
6
solvent.
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Soak a cleaning brush in the solvent and scrub the inside
7
of the inlet weldment. Repeat 10 times.
Rinse the inlet with the solvent.
8
"Consumables and Parts for
on page 44.)
and wait for the GC
"To Change the Liner and
on page 56.)
"To Replace
on page 59.)
Maintaining Your GC