Maintaining Your GC
Score the column using a glass scribing tool. The score
4
must be square to ensure a clean break.
5
Break off the column end by supporting it against the
column cutter opposite the scribe. Inspect the end with a
magnifying loupe to make certain that there are no burrs
or jagged edges.
6
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
Insert the column into the detector until it bottoms.
7
Slide the column nut and ferrules up the column to the
8
detector and finger- tighten the nut.
9
Pull out 1 mm of column. Tighten the nut an additional
1/4 turn with a wrench or until the column does not
move.
Maintaining the TCD
Bad
Good
12
235