Maintaining Your GC
Lower the collector assembly into the housing.
17
Insert the three screws and tighten (to 18 inch- pounds).
18
19
Connect the ignitor extension cable.
20
Verify assembly:
a
Check the FID leakage current. (See
Leakage Current"
Bakeout the detector. (See
b
page 229.)
Check the FID baseline. (See
c
Baseline"
on page 225.)
Reset the EMF counters. See
21
the Operation Manual.
Maintaining the FID
"To Check the FID
on page 224.)
"To Bakeout the FID"
"To Check the FID
To Reset an EMF Counter
11
on
in
223