EITHER
The washer-disinfector must comply with ISO 15883, must have suitable baskets to hold small products and must have
rinsing connections with diameter of approximately 16mm for the attachment to product lumen.
Correctly place the product into a suitable rack, connect all lumens to the rinsing connectors and start the automated cleaning.
Instructions of the manufacturer of the washer disinfector must also be observed.
Carefully follow the instructions provided by the disinfection solution manufacturer.
The following automated process
2 min
5 min
3 min
2 min
3 min
20 min (at least) Dry at 100°C.
OR
02

MANUAL CLEAN +

The following validated EMS process can be used with and without ultrasound.
DO NOT use any ultrasound bath cleaning procedure with the handpieces: it may destroy the products. It is ONLY intended
to be used with EMS PIEZON
and sonicate 10 minutes.
Validated
product
0.5%
neodisher
ultrasound
MediClean
(Dr. Weigert)
0.5%
neodisher
ultrasound
MediClean
(Dr. Weigert)
Then disinfect:
ASP CIDEX
®
warnings and instructions for use.
Immerse the product completely into CIDEX
5 min
with disinfectant solution (use a syringe if needed).
The ASP CIDEX
1 min
volume
(each rinse)
Do not reuse the water for rinsing or any other purpose. Residues of disinfectant may cause serious side effects.
And dry
At room temperature (15-25°C), use an air pistol (compressed air, max. 3 bars) to fully dry the lumen and the whole product until
no more residues of water are present (visible or detectable).
05
INSPECT
If stains are still visible on the product after cleaning/disinfection, the entire cleaning/disinfection procedure must be
repeated. Products with visible damage, chip/flake loss, corrosion or deformation must be disposed of (no further use is
allowed). Check also the integrity of O-rings and gaskets and replace if damaged or deformed.
Verify the product to be fully dry. In case of detection of residues of water, remove these using an air pistol (clean compressed
air). Fully dry the lumen and the whole product until no more residues of water are present (visible or detectable).
3 For example: Miele Professional G 7836 CD with Miele Rack E429
4 cold tap water = 16°C +/- 2°C
5 cold demineralized water = 20°C +/- 2°C
6
6 For example: Bandelin, Sonorex 1028K, 35kHz
7 e.g. 2 gallons
02 03 04
AUTOMATIC CLEAN, DISINFECT AND DRY WITH WASHER DISINFECTOR
can be used to achieve an A0 Level of 3000:
3
Pre-clean with cold tap water
Drain
Clean at 55°C with tap water and 0.5% neodisher MediClean Dental (Dr Weigert) solution.
Drain
Rinse and neutralize with cold demineralized water
Drain
Rinse with cold demineralized water
Drain
Perform thermal disinfection (final rinse) with deionized water at 93°C minimum.
Drain
03
Instruments and file holders. In case of ultrasound bath cleaning, place products in a metal sieve
®
EMS products
Process
to clean/
disinfect
without
all EMS
products
bath
EMS PIEZON
®
with
Instruments
and file
bath
6
holders
OPA solution must be used undiluted, within its useful and shelf life, and in compliance with the manufacturer's
OPA disinfectant requires a total of three rinses : keep the product totally immersed, and use a large
®
of fresh water.
7
.
4
.
5
.
5
DISINFECTION +
Brush the devices in a cleaning solution of 0.5% neodisher in deionized water at 40°C
with a suitable soft bristled brush until all visible residue has been removed.
For products with lumens, flush all lumens with a spray gun (water jet
15 s
gun, with a static water pressure of 2 bar) with cold tap water.
Place the product in 0.5% solution of neodisher with deionized water at
15 min
40°C. Make certain that all lumens are filled with cleaning solution (use
(10 min with
a syringe if needed).
ultrasound)
Rinse all lumens by flushing with a spray gun (water jet gun, with a static
15 s
water pressure of 2 bar) with cold deionized water
Rinse the whole product under cold running deionized water.
10 s
At room temperature (15-25°C), use an air pistol (compressed air, max. 3 bars) to
fully dry the lumen and the whole product until no more residues of water are present
(visible or detectable).
OPA solution at minimum 20°C. Make certain that all lumens are filled
®