Horizontal Rotation of the Receptor Assembly is not allowed.
The Receptor Assembly may be damaged if it bumps the
Column. From the Horizontal position of the Receptor, tilt the
o
Assembly a minimum of 45
. Rotate the Assembly and
o
position the Assembly at 0
.
Attention, never do place hands or fingers neither let patient
hands or fingers inside the Tilting Assembly as shown below:
it may cause serious injuries to patient or operator.
As well, make sure that the patient extremities are inside the
tabletop or accessories limits during operation: movement of
parts may cause serious damages to patient.
0232F EN 20191213