DX-D 400
User Manual
Horizontal Rotation of the Receptor Assembly is not allowed.
The Receptor Assembly may be damaged if it bumps the
Column. From the Horizontal position of the Receptor, tilt the
o
Assembly a minimum of 45
. Rotate the Assembly and
o
position the Assembly at 0
.
Attention, never do place hands or fingers neither let patient
hands or fingers inside the Tilting Assembly as shown below:
it may cause serious injuries to patient or operator.
As well, make sure that the patient extremities are inside the
tabletop or accessories limits during operation: movement of
parts may cause serious damages to patient.
0232F EN 20191213
77