Table 5-1
Error Codes
ERROR
DESCRIPTION
System failure. This indication may appear together with an
-- -- -- -- -- --
error on the Console, and indicates that the error is not
on Display
correctable unless the equipment is turned OFF.
E01, E02
Communication error.
E03
System failure.
The X-ray Generator has activated "Preparation" without a
E04
Console command intervention.
E05
External exposure activated during power-up.
"Exposure" and/or "Preparation" orders are activated during
E06
power-up.
E07, E08
X-ray Tube configuration error.
Generator Overload error. The exposure has been
interrupted because during exposure has been produced
arcing or bad function on the HV circuitry (X-ray Tube, HV
Transformer and/or HV Cables) or a failure of IGBT module
E09
(overheated or defective IGBTs) has been detected.
It can be also shown making a high power and long
exposure with the X-ray tube cool (X-ray Tube has not been
warmed-up).
E10, E11
System failure.
E12
No mA during exposure or mA value is out of range.
E13
No kV during exposure or kV value is out of range.
E14, E15
System failure.
E16
Invalid value of: kV, mA or kW.
E17
Communication error or system failure.
Rotor error. The X-ray tube anode is not rotating while
E18
"Prep" is active, then exposures are inhibited, or the X-ray
tube anode is rotating without console command.
0232F EN 20191213
WHAT TO DO
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Turn the Generator OFF, check the proper external cable connections
and then turn the Generator ON. If the equipment remains inoperative,
turn it OFF and call Field Service.
Turn the Generator OFF and ON.
T
th G
t
OFF
d ON
If the equipment remains inoperative, turn it OFF and call Field Service.
Release any external exposure device or buttons.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Release all the controls.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
It is not necessary to touch the "Reset" button, the error indication will
disappear automatically.
If the error code persists, turn the Generator OFF and wait 30 minutes
before turning it ON again.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
Repeat with same technique values, If the error code persists try with
Repeat with same technique values, If the error code persists try with
another combinations of kV and mA values.
If the equipment remains inoperative, turn it OFF and call Field Service.
Touch the "Reset" button.
If the error code persists, turn the Generator OFF and ON again.
If the equipment remains inoperative, turn it OFF and call Field Service.
Decrease kV, mA or both. touch the "Reset" button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
It is not necessary to touch the "Reset" button, the error indication will
disappear automatically.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
DX-D 400
User Manual
119