GUIDANCE AND MANUFACTURER'S DECLARATION - - ELECTROMAGNETIC IMMUNITY
The DX- -D400 radiographic room is intended for use in the electromagnetic environment specified below.
The customer or the user of The DX- -D400 radiographic room should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4 (*)
Surge
IEC 61000--4--5 (**)
Voltage
dips,
short
interruptions
and
voltage
variations on power supply
input lines.
IEC 61000--4--11
Power frequency (50 Hz)
magnetic field
IEC 61000--4--8
NOTE - - U
is the a.c. mains voltage prior to application of the test level.
T
32
(IEC 60601-1-2:2007)
IEC 60601-1-2:2007
Test level
6 kV contact
¦
8 kV air
¦
2 kV for power supply lines
¦
1 kV for input/output lines
¦
1 kV line(s) to line(s)
¦
2 kV line(s) to earth
¦
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
< 5% U
T
(>95% dip in U
)
T
for 250 cycles
< 5% U
T
(>95% dip in U
)
T
for 0.5 cycle
< 5% U
T
(>95% dip in U
)
T
for 1 cycle
3 A/m
Compliance
Electromagnetic environment - - guidance
level
6 kV
Floors should be wood, concrete or ceramic
¦
tile. If floors are covered with synthetic material,
8 kV
the relative humidity should be at least 30%.
¦
2 kV
¦
Mains power quality should be that of a typical
commercial or hospital environment.
N/A
1 kV
¦
Mains power quality should be that of a typical
commercial or hospital environment.
2 kV
¦
60%
for 5 periods
30%
for 25 periods
Mains power quality should be that of a typical
commercial or hospital environment. If the user
>95%
of the Radiographic Room requires continued
for 250 periods
operation during power mains interruptions, it is
recommended that the Radiographic Room be
powered from an uninterruptible power supply
or a battery.
>95%
for 0.5 periods
>95%
for 1 period
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
3 A/m
typical commercial or hospital environment.